Universal Robot UR5 arm, PV010204 subcomponent of Aesculap Aeos Digital Microscope System, PV010
- Company
- Aesculap Implant Systems Llc
- Recall Initiated
- February 19, 2021
- Recall Number
- Z-1409-2022
- Quantity
- 1
- Firm Location
- Center Valley, PA
Reason for Recall
Malfunction of the robotic arm preventing further movement of the robotic arm and requiring restart of the robotic digital microscope system.
Distribution
United States: NE
Lot / Code Info
UDI-DI: 04046955206742, Serial Number: SN1004
Root Cause
Material/Component Contamination
Action Taken
An Urgent Market Withdrawal Notification letter issued February 19, 2021 was mailed USPS certified. The letter identifies the Universal Robot UR5 arm as part of hte Aesculap Aeos System and requests the customer return the acknowledgement form and schedule repair by service technician.