RecallDepth

Sara Plus Active Floor Lift

This recall has been terminated (originally issued May 21, 2022).

Company
Arjohuntleigh Polska Sp. Z.O.O. Ul. Ks. Piotra Wawrzyniaka 2 Komorniki Poland
Recall Initiated
April 5, 2022
Posted
May 21, 2022
Terminated
September 5, 2024
Recall Number
Z-1098-2022
Quantity
4449 devices
Official Source
View on FDA website ↗

Reason for Recall

The device may emit smoke or ignite.

Distribution

US Nationwide and Worldwide distribution

Lot / Code Info

Model Numbers: HEP0001, HEP0001-AU, HEP0001-BR, HEP0001-CN, HEP0001-JP, HEP0001-UK, HEP0001-US, HEP1001, HEP1001-AU, HEP1001CON4869, HEP1001-US, HEP2001, HEP2001-BR, HEP2001-CN, HEP2001-UK

Root Cause

Device Design

Action Taken

A customer letter was sent to consignees on 04/08/2022. The customer are asked to acknowledge the notification, complete and sign off an enclosed Customer Response Form and return it to the local Arjo office (address indicated in the notification). The notification contains the list of precautionary measures to be taken by caregivers in order to minimize the likelihood of the failure until their lifts are corrected by Arjo. The device correction (including PCB upgrade and cover replacement) is to be performed by Arjo personnel, free of charge.