RecallDepth

Nimbus PainPro Ambulatory Infusion Pumps with software version v4, Model Number IT1072, sold as kit IT1075-DUKE-HSP, IT1076-DUKE-HSP and IT1077-DUKE-HSP

Company
Infutronix Llc
Recall Initiated
October 19, 2021
Recall Number
Z-0832-2022
Quantity
74 units
Firm Location
Natick, MA

Reason for Recall

An unreleased Version 4 Protocol Library was installed on Nimbus PainPro (H) infusion pumps in error that may impact patient safety if used on patients.

Distribution

US Nationwide distribution in the states of NC and OH.

Lot / Code Info

(UDI): (01)00817170020086; Lot: 2009231440; Serial Numbers: 807483, 807484, 807485, 807486, 807487, 807489, 807490, 807491, 807492, 807493, 807494, 807495, 807496, 807497, 807498, 807499, 807500, 807502, 807651, 807652, 807653, 807507, 807508, 807509, 807510, 807511, 807512, 807513, 807514, 807515, 807516, 807517, 807518, 807519, 807520, 807521, 807522, 807523, 807524, 807525, 807526, 807527, 807528, 807529, 807530, 807531, 807532, 807533, 807534, 807535, 807536, 807537, 807538, 807539, 807540, 807541, 807542, 807543, 807544, 807545, 807546, 807547, 807548, 807549, 807550, 807551, 807552, 807553, 807554, 807654, 807655, 807656, 807657, 807658

Root Cause

Employee error

Action Taken

An IMPORTANT PRODUCT INFORMATION letter dated 10/19/21 was sent to customers. InfuTronix LLC has recently become aware of an unreleased Version 4 Protocol Library for Duke University Hospital was installed on your Nimbus PainPro (H) infusion pumps in error that may impact patient safety if used on patients. Please ensure that all your potential customers and potential users in your facility are made aware of this notification and the recommended actions. Product Issue Version 4 of the Duke Protocol Library was installed on your infusion pumps in error. Affected Product Details Nimbus PainPro (H) Pump, Model numbers: IT1072, sold as kits IT1075-DUKE-HSP Total number of infusion pumps affected: 3 Lot numbers involved: 2009231440 Instructions 1. Place all units you have in-house on product-hold and stop further distribution 2. Inform your customer to discontinue use of the affected infusion pumps. Product Correction InfuTronix will be contacting you to resolve this issue as quickly as possible. Contact Information Your Account Manager, Vann Griffin, email: vann.griffin@nimbuspainpro.com, phone: (919) 606-8339 Or John LaFratta, Vice President of Sales, email: john.lafratta@ nimbuspainpro.com , phone: 315-729-5564. Please be assured that maintaining a high level of safety and quality is our highest priority. If you have any questions, please contact us immediately.

More recalls by Infutronix Llc