RecallDepth

MPR Slotted Mallet, labeling etched onto devices as follows: Medical Products Resource 800.524.5194 www.m-p-r.com MPR-1120-1000 (01)00840001405868

Company
Implant Resource Inc
Recall Initiated
July 21, 2021
Recall Number
Z-0122-2022
Quantity
6 units
Firm Location
Eagan, MN

Reason for Recall

Potential for mallet head to become separated from mallet handle. There is also significant deformations on striking surface that were not expected

Distribution

US distribution to Florida and New Hampshire

Lot / Code Info

Model Number: MPR-1120-1000, Lot Number: 200911-1-1

Root Cause

Device Design

Action Taken

Customers were initially contacted via email on 07/21/2021 to initiate the return of affected devices. A formal letter was sent to customers on 09/03/2021 via UPS. Instructions included to review the notification and ensure affected personnel are notified of the recall, immediately locate and quarantine any affected product in inventory, and to immediately return any affected product to Medical Products Resource.