MPR Slotted Mallet, labeling etched onto devices as follows: Medical Products Resource 800.524.5194 www.m-p-r.com MPR-1120-1000 (01)00840001405868
- Company
- Implant Resource Inc
- Recall Initiated
- July 21, 2021
- Recall Number
- Z-0122-2022
- Quantity
- 6 units
- Firm Location
- Eagan, MN
Reason for Recall
Potential for mallet head to become separated from mallet handle. There is also significant deformations on striking surface that were not expected
Distribution
US distribution to Florida and New Hampshire
Lot / Code Info
Model Number: MPR-1120-1000, Lot Number: 200911-1-1
Root Cause
Device Design
Action Taken
Customers were initially contacted via email on 07/21/2021 to initiate the return of affected devices. A formal letter was sent to customers on 09/03/2021 via UPS. Instructions included to review the notification and ensure affected personnel are notified of the recall, immediately locate and quarantine any affected product in inventory, and to immediately return any affected product to Medical Products Resource.