RecallDepth

Monoject 0.9% Sodium Chloride Flush Syringe, 3 mL Fill, STERILE, Product Code 8881570123 The proposed device is indicated for use in flushing compatible intravenous tubing systems and indwelling intravascular access devices.

Company
Cardinal Health
Recall Initiated
August 4, 2021
Recall Number
Z-2285-2021
Quantity
267,217,860 eaches in total
Firm Location
Mansfield, MA

Reason for Recall

Potential for the plunger to draw back after the air has been expelled and reintroduce air back into the syringe.

Distribution

US Nationwide Distribution: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV and WY.

Lot / Code Info

20I1524 20D0694 20I1584

Root Cause

Device Design

Action Taken

On August 4, 2021 Cardinal Health issued URGENT MEDICAL DEVICE RECALL notices and response forms to customers via overnight delivery. Customers are advised to take the following actions: 1) REVIEW your inventory for affected product. 2) SEGREGATE and QUARANTINE all on-hand product per the Affected Product table. 3) COMMUNICATE the notice to clinical staff through safety huddles to ensure awareness of this product issue. 4) RETURN the enclosed acknowledgment form via fax to 614-652-9648 or email to gmbfieldcorrectiveaction@cardinalhealth.com, whether or not you have affected product. 5) CONTACT the appropriate Customer Service group to arrange for return and credit of any affected product Monday Friday between 8:00am - 5:00pm EST: " Hospital800-964-5227 " Federal Government800-444-1166 " Distributor800-635-6021 " All other Customers888-444-5440 6) CUSTOMERS that did not receive product directly from Cardinal Health should contact the location where they purchased it. For questions related to the notification and/or acknowledgement form that are not adequately addressed in this letter, please contact the market action team at: GMB-FieldCorrectiveAction@cardinalhealth.com or call 800-292-9332.

More recalls by Cardinal Health

View all recalls by this company →