AMX Navigate Mobile X-Ray System

This recall has been terminated (originally issued September 29, 2021).

Company
Ge Healthcare, Llc
Recall Initiated
June 24, 2021
Posted
September 29, 2021
Terminated
June 14, 2023
Recall Number
Z-2590-2021
Quantity
3 units
Firm Location
Waukesha, WI
Official Source
View on FDA website ↗

Reason for Recall

An unexpected column motion while attempting to park the column on the AMX Navigate mobile X-ray system.

Distribution

Domestic Distribution Only

Lot / Code Info

AMXNV2000001WK, AMXNV2000002WK, DF2401600003WK

Root Cause

Process control

Action Taken

On June 24, 2021 the firm contacted the facilities that received the device via email. An AMX Navigate system was provided to 3 healthcare facilities for evaluation under an agreement with GE Healthcare. While at the healthcare facilities, these systems remained under the ownership of GE Healthcare. After one of the healthcare facilities alerted GE Healthcare to a potential issue with its system, GE Healthcare removed the system from the 3 healthcare facilities in order to perform a detailed engineering analysis as part of its root cause investigation. These 3 systems have not been returned to the healthcare facilities and remain in GE Healthcare s possession.