Computed Tomography X-ray System; Model uCT 530; Rx; UDI: (01) GTIN: 06971576831012 - Product Usage: intended to produce cross-sectional images of the body using computer reconstruction of x-ray transmission data taken at different angles and planes and indicated for the whole body (including head, neck, cardiac, and vascular).

Company
Shanghai United Imaging Healthcare Co., Ltd. No. 2258 Chengbei Rd, Jiading Ind. Shanghai China
Recall Initiated
June 20, 2020
Posted
July 27, 2020
Recall Number
Z-2704-2020
Quantity
4

Reason for Recall

The metal edge overlap of the mylar strip in the gantry was uneven, which may create a sharp edge.

Distribution

US Nationwide distribution including in the states of CA, IN and TX. See uploaded distribution list for more details.

Lot / Code Info

SN 300162, 300163, 300174, 300176

Root Cause

Device Design

Action Taken

A Customer Notification Letter was sent to the affected consignees on 06/20/2020 via email. United Imaging will replace the mylar strip with an improved one on the affected products via on-site update. On 8/30/2024, a correction notice was sent identifying an additional serial number.