RecallDepth

Positron Emission Tomography and Computed Tomography System, Model: uMI 550, Material Number: 88000057 - Product Usage: This system is intended to be operated by qualified healthcare professionals to assist in the detection, localization, diagnosis, staging, restaging, treatment planning and treatment response evaluation for diseases and disorders in, but not limit to, oncology, cardiology and neurology.

Company
Shanghai United Imaging Healthcare Co., Ltd. No. 2258 Chengbei Rd, Jiading Ind. Shanghai China
Recall Initiated
March 26, 2020
Terminated
August 2, 2023
Recall Number
Z-2819-2020
Quantity
3

Reason for Recall

In affected software version, of Positron emission tomography and computed tomography system, the CT Hounsfield unit curve may not display in real time when using the bolus tracking protocol with the tracker scan function. If this occurs, the subsequent clinical scanning protocols will need to be started manually, which may effect image enhancement, and may result in patient rescan.

Distribution

US Nationwide distribution including in the state of TX.

Lot / Code Info

UDI: 06971576832026, Software Version: R001.3.0.0.750505, Serial Numbers: 200017, 200023, 200024

Root Cause

Software design

Action Taken

On March 26, 2020, Customer Notification Letters were emailed to customers. The recalling firm has released a software hotfix to address the issue for the affected products. The corrective action of on-site upgrade will be provided free of charge. For further questions, feel free to contact the Customer Contact Center at (855) 221-1552 (toll free) or (832) 699-2799 (direct).

More recalls by Shanghai United Imaging Healthcare Co., Ltd. No. 2258 Chengbei Rd, Jiading Ind. Shanghai China

View all recalls by this company →