RecallDepth

LivaNova USA Sterile Disposable Connector,1/2x1/2, 20 units per case

This recall has been terminated (originally issued November 18, 2019).

Company
Livanova Usa Inc.
Recall Initiated
November 18, 2019
Terminated
August 19, 2021
Recall Number
Z-2642-2020
Quantity
140 units
Firm Location
Arvada, CO
Official Source
View on FDA website ↗

Reason for Recall

Some packages of sterile disposable connectors contained a 3/8x1/4 reducer instead of the specified 1/2x1/2 connector. No significant risk to health is expected as the issue is easily detectable and the different shape would prevent use of the connector.

Distribution

Distributed to medical facilities in 5 states including: Minnesota, Kansas, Missouri, Utah and Washington.

Lot / Code Info

Item# EC2135S, Lot No. 1923900091

Root Cause

Packaging process control

Action Taken

Affected customers were contacted by letter sent via email or by phone call on November 18th, 2019. Customers were informed that a single lot of LivaNova USA Sterile Disposable Connector item# EC2135S, lot number 1923900091, contained a 3/8 x 1/4 reducer instead of the specified 1/2 x 1/2 connector. Customers were requested to check their inventory, quarantine (and do not use) any affected devices, complete the Customer Response Form and submit the completed form by email to USFSN@livanova.com to confirm quantity of affected products in the facility. Customers were informed that affected product should be returned to LivaNova, and that their LivaNova reference person would contact customers to organize the replacement of the affected products. Customers were informed that no patient injuries have been reported and that the event is unlikely to cause any harm to the patient or user. Customers were requested to ensure that this Medical Device Correction notice is communicated to all personnel within their organization who need to be aware of it, and, if a customer has transferred any of the affected devices to a third party, to please communicate this information to them and inform the LivaNova Quality Assurance Team at USFSN@livanova.com. Customers with questions may contact LivaNova at (800) 986-4702 or by e-mail to USFSN@livanova.com. Customers were informed that this action is being reported to the Food and Drug Administration (FDA). Adverse reactions or quality problems experienced with the use of this product may be reported to LivaNova at customerquality@livanova.com or the FDAs MedWatch Adverse Event Reporting program either online, by regular mail or by fax.

More recalls by Livanova Usa Inc.

View all recalls by this company →