DABRA RA-308 Excimer Laser, SN: ******* - Product Usage: The device is intended for use in ablating a channel in occlusive peripheral vascular disease.

This recall has been terminated (originally issued July 24, 2020).

Company
Ra Medical Systems, Inc.
Recall Initiated
July 24, 2020
Terminated
August 9, 2023
Recall Number
Z-0050-2021
Quantity
54 units
Firm Location
Carlsbad, CA
Official Source
View on FDA website ↗

Reason for Recall

Due to the taller wheel mounting on the laser, the laser may fall over potentially causing user or patient injury.

Distribution

Worldwide distribution - US Nationwide distribution including in the states of AL, AZ, CA, FL, ID, IL, IN, KS, KY, LA, MA, MD, MI, NC, NV, NY, OR, OK, PA, TX, WA and the country of Japan.

Lot / Code Info

Model Number DABRA Excimer Laser RA-308; Serial Numbers: RA00027, RA00039, RA00046, RA00049, RA00050, RA00054, RA00055, RA00056, RA00057, RA00058, RA00063, RA00064, RA00065, RA00067, RA00069, RA00070, RA00073, RA00074, RA00081, RA00082, RA00083, RA00084, RA00085, RA00086, RA00088, RA00089, RA00097, RA00098, RA00103, RA00114, RA00115, RA00117, RA00118, RA00120, RA00121, RA00124, RA00125, RA00127, RA00135, RA00136, RA00137, RA00138, RA00143, RA00147, RA00148, RA00153, RA00154, RA00160, RA00164, RA00165, RA00167, RA00170, RA00175, and RA00177.

Root Cause

Device Design

Action Taken

On August 10, 2020, the firm sent an "Urgent Medical Correction" letter to customers, informing them of the incorrect laser wheel size, which upon moving or adjusting the laser with the wheels in locked position may fall over. If this were to happen during a medical procedure, user or patient injury may occur.