Irix-C Forceps Inserter; Product Code/UDI: T066-0059/ M697T06600591

This recall has been terminated (originally issued May 12, 2020).

Company
Xtant Medical
Recall Initiated
May 12, 2020
Terminated
March 19, 2021
Recall Number
Z-2500-2020
Quantity
4 units
Firm Location
Belgrade, MT
Official Source
View on FDA website ↗

Reason for Recall

The strike plate fractured off from the forceps inserter devices with minimal impaction or force due to unsatisfactory weld.

Distribution

AZ, TX and WV.

Lot / Code Info

LOT: XS1920458A

Root Cause

Process design

Action Taken

On May 12, 2020 a customer letter (in an envelope marked URGENT) was Federal Expressed priority overnight-signature required to the three customers who received the affected devices. On May 19, 2020, an additional customer letter (in an envelope marked URGENT) was Federal Expressed priority overnight-signature required to one of the three affected customers.