Medtronic RestoreSensor, Model 37714; and, RestoreSensor SureScan, (MRI), Model 97714, Implantable Neurostimulators. Used to treat chronic, intractable pain of the trunk and/or limbs-including unilateral or bilateral pain.

This recall has been terminated (originally issued June 7, 2016).

Company
Medtronic Neuromodulation
Recall Initiated
May 3, 2016
Posted
June 7, 2016
Terminated
August 4, 2017
Recall Number
Z-1926-2016
Quantity
60,941
Firm Location
Minneapolis, MN
Official Source
View on FDA website ↗

Reason for Recall

Loss of therapy during recharging of device. Reportedly, a charging session was terminated prior to obtaining recharge threshold voltage of 3.615, which triggered rapid battery depletion state. As a result, the device depleted to 1.925 volts in one to two days rather than the typical 30 days. Insufficient coupling between recharger and device during recharge session was found to be key factor.

Distribution

Worldwide Distribution

Lot / Code Info

All lot/serial numbers are affected for both Model numbers 37714 and 97714.

Root Cause

Under Investigation by firm

Action Taken

Consignees were sent on 5/3/2016 a Medtronic "Urgent Medical Device Safety notification" letter dated May 2016. The letter was addressed to "Healthcare Professional". The letter described the problem and the product involved in the recall. The letter listed the recommendation to patients to follow current recharge instructions, paying particular attention to Charging Efficiency and Battery Charge Level indicators on the recharger. Consignees were also requested to complete and return the "Physician Reply Form:. For questions contact Medtronic Neuromodulation Technical services at 1-800-707-0933 weekdays 7 am - 6 pm CST.