Philips Healthcare PIIC Classic Upgrade, 866117 Physiological, Patient Monitor (With Arrhythmia Detection or Alarm)

This recall has been terminated (originally issued February 23, 2016).

Company
Philips Electronics North America Corporation
Recall Initiated
November 25, 2015
Posted
February 23, 2016
Terminated
September 6, 2017
Recall Number
Z-0857-2016
Quantity
5569 units
Firm Location
Andover, MA
Official Source
View on FDA website ↗

Reason for Recall

Reconstructed ECG leads viewed or printed at the Information Center iX may misrepresent the ECG waveform in specific leads.

Distribution

Worldwide Distribution-US (nationwide) and the countries of Algeria Australia Austria Belgium Bulgaria Canada Denmark Estonia Finland France Germany Hong Kong Hungary Iceland India Indonesia Iraq Ireland Israel Italy Japan Kenya Korea Kuwait Luxembourg Malaysia Mongolia Morocco Netherlands New Zealand Nigeria Norway Poland Qatar Romania Russian Federation Saudi Arabia Singapore Slovenia South Africa Spain Sweden Switzerland Taiwan Turkey United Arab Emirates and United Kingdom.

Lot / Code Info

Software B.O

Root Cause

Software Design Change

Action Taken

Philips Healthcare issued a Field Safety Notice on 11/25/15 to inform customers of the issue, identifies details of the units affected, gives instructions on actions to be taken by the customer and identifies what action Philips plans to take to remedy the issue.Philips will provide a software correction free of charge. On 1/28/16 the Philips issued an updated letter to users.Philips has initiated a correction to address this issue. A software correction will be provided to customers with impacted devices at no charge. Philips is asking customers to follow the Action to be Taken by Customer/User section of the Field Safety Notice: Arrhythmia alarms are not affected. ST alarms, ST values and measurements are also not affected and could be used to assist with the decision to obtain a diagnostic ECG. The ECG waves at the bedside are not affected. Printing ECG waves from the bedside is an alternative to printing from the PIIC iX. If you need any further information or support concerning this issue, contact your local Philips Healthcare Customer Service representative or the Customer Care Solutions Center at (800) 722-9377 with questions or concerns about this Device Correction.