Smith & Nephew MOSAICPLASTY DP- Disposable Harvesting System 8.5 mm, Sterile Part Number: 7209237 Intended Use: For arthroscopic (knee and ankle) and open osteochondral grafting fix for osteochondral (cartilage) defects.

Company
Smith & Nephew, Inc., Endoscopy Div.
Recall Initiated
September 10, 2015
Posted
October 22, 2015
Terminated
June 23, 2021
Recall Number
Z-0158-2016
Quantity
1246 units
Firm Location
Mansfield, MA

Reason for Recall

Sterility of device is compromised due to breach in the packaging

Distribution

Worldwide Distribution - US Nationwide, Canada, Austria, Australia, Belgium, China, Spain, Finland, France, Great Britain, Greece, Germany, Hungary, India, Italy, Japan, Korea, Lithuania, Latvia, Netherlands, Poland, Portugal, Sweden, Thailand, Turkey, Ukraine and Venezuela

Lot / Code Info

50069280 50214249 50260936 50334160 50380067 50410625 50442600 50084563 50217938 50266144 50337465 50383809 50412120 50448621 50097669 50227860 50276674 50352665 50385382 50415177 50451051 50112174 50231228 50282533 50353524 50389835 50416603 50451744 50117989 50236270 50310331 50356943 50391478 50424410 50452069 50131406 50242373 50317889 50362696 50396029 50428473 50455380 50139181 50246739 50320168 50370396 50399036 50428686 50457280 50148410 50254144 50324084 50372214 50401621 50432549 50461109 50199543 50254145 50328908 50374300 50404803 50436553 50469462 50203454 50258249 50332285 50378312 50407329 50436675 50475144 50481307

Root Cause

Package design/selection

Action Taken

Smith & Nephew issued Recall Notification Letter dated September 10, 2015 via overnight Federal Express. Customers were requested to inspect inventory and locate any unused devices from the above listed product and batch numbers, and quarantine them immediately. Complete the Inventory Return Certification Form on the following page, indicating the lot numbers and quantities that need to be returned and include your phone and fax number in the spaces provided. If you do not have product to return, please place an x in the column No product to return. Send the completed form to the Smith & Nephew's Field Action Department via email at FieldActions@smithnephew. com or fax +1-901-566-7975 to obtain a return authorization (RA) number. Return any affected product to the address listed on the Inventory Return Certification Form. Please indicate the RA number on your return shipment.