RecallDepth

Hypodermic Needle Pro Edge" Safety Device with Syringe 20G x1 10mL Luer lock syringe; Reorder Number: 4102010 Product Usage: The Needle Pro Edge Safety Hypodermic Needle is intended for injection or aspiration of fluids utilizing a luer lock or luer slip syringe.

This recall has been completed (originally issued December 30, 2014).

Company
Smiths Medical Asd, Inc.
Recall Initiated
November 21, 2014
Posted
December 30, 2014
Recall Number
Z-0898-2015
Quantity
47,200
Firm Location
Keene, NH
Official Source
View on FDA website ↗

Reason for Recall

Needle is not captured in the needle safety sheath in specific lots.

Distribution

US Nationwide Distribution in the states of AZ CA FL GA IL IN LA MA MN MO MS MT NC ND NJ NV NY OH OR PA RI TN TX VA WA WI WV

Lot / Code Info

Lot Numbers: 2756526, 2771498

Root Cause

Process control

Action Taken

Smith Medical sent on 11/24/2014, consignees an Urgent Field Safety Notice via FedEx Overnight. The notification instructs customers to inspect their inventory and complete and return the attached Urgent Field Safety Notice Confirmation Form to coordinate return of affected product to Smiths Medical for replacement/credit. For questions regarding this information, please contact Smiths Medicals Customer Service Department at 800-258-5361.

More recalls by Smiths Medical Asd, Inc.

View all recalls by this company →