CARESTREAM DIRECTVIEW CR Software Generates digital mammographic images that can be used for screening and diagnosis of breast cancer.

Company
Carestream Health, Inc.
Recall Initiated
October 30, 2014
Posted
December 17, 2014
Terminated
March 10, 2017
Recall Number
Z-0820-2015
Quantity
59 units
Firm Location
Rochester, NY

Reason for Recall

Reduced mammographic image quality when attempting to print true size multi-format images

Distribution

Worldwide Distribution: US distribution to states of : NY, OH, and NC; and countries of: Argentina, Brazil, China, Chile, Colombia, France, Greece, Hong Kong, India, Japan, Korea, Mexico, Peru, Spain, Taiwan, and Thailand.

Lot / Code Info

Software Verions: 1) 5.7.312.27 , 2) 5.7.312.27, 3) 5.7.312.2031

Root Cause

Software design

Action Taken

The firm, Carestream Health Inc., sent an "URGENT: Medical Device Recall" notice dated October 28, 2014 to its customers. The notice described the product, problem and actions to be taken. The customers were instructed to use the one-up printing option rather the multi-format printing option if a hard copy is required. Carestream has a software update that will address this issue. A Carestream service representative will contact all customers to schedule a visit to correct this software issue. If you have any questions or concerns or would like to request the installation of the previous software version that does not have this issue, please contact the Carestream Customer Care Center in the U.S. at 1-800-328-2910, available 7 days per week on a 24 hour basis. Outside of the U.S. please call your local service support number.