RecallDepth

Rusch, Soft Rubber (Nelaton and Nelaton Robinson), Bladder Catheter, Sterile. These catheters are indicated for the routine drainage of urine from the bladder. These catheters are not intended or designed for indwelling use.

This recall has been terminated (originally issued September 18, 2014).

Company
Teleflex Medical
Recall Initiated
September 4, 2014
Posted
September 18, 2014
Terminated
February 22, 2017
Recall Number
Z-2661-2014
Quantity
193,363 ea. + 1,140 units (total 194,503 units)
Firm Location
Research Triangle Park, NC
Official Source
View on FDA website ↗

Reason for Recall

Misbranding: Although the product labeling identifies the catheters as "soft rubber," the Natural Rubber Latex caution statement was not printed on the product label.

Distribution

Worldwide Distribution-USA (nationwide) and the countries of Austria, Belgium, Bahrain, Belgrad, Canada, Croatia, Cypress, Czeck Republic, Denmark, Finland, Germany, Great Britain, Hungary, Ireland, Italy, Libya, Netherlands, Norway, Romania, Russia, Saudi Arabia, Slavonia, Slovakia, Spain, Switzerland, Turkey, Portugal and Zambia.

Lot / Code Info

Product Code #402100060, Lot #13491 & 13501; Product Code #402140120, Lot #13371, 13391 & 13511; Product Code #401140140, 13501, 14091, 14141, 13271, 13461, 13481, 14061, 14081 & 14101; Product Code #402140160, Lot #13411, 13291, 13311, 13301, 13391 & 13491; Product Code #402140240, Lot #13311 & 13391; Product Code #403300100, Lot #13311, 13411, 13461, 13481, 14041, 14091 & 14131; Product Code #403340140, Lot #13291; Product Code #403900080, Lot #13371 & 13411; Product Code #403900100, Lot #13261 & 13441; Product Code #403900120, Lot #13351, 13451, 13511 & 13461 and Product Code #410200200, Lot #13411. On 1/9/2015 recall expanded to include: Product Code: 40214-000080, Lot numbers 14101 & 13301; Product Code: 402140-000140, Lot number: 14281; Product Code: 402140-0001160, Lot number: 13491; Product Code: 402140-000180, Lot number: 13391 and Product Code; 402140-000240, Lot numbers: 13391 & 14027.

Root Cause

Labeling Change Control

Action Taken

Consignees were notified via letter on/about 09/04/2014. Subsequently, consignees were notified of the expanded recall on 01/13/2015.