Codman Certas - In Line Valve Only; Product Code: 82-8800 The Codman Certas Programmable valve is an implantable device that provides constant intraventricular pressure and drainage of CSF (cerebrospinal fluid) for the management of hydrocephalus.

This recall has been terminated (originally issued August 12, 2014).

Company
Codman & Shurtleff, Inc.
Recall Initiated
July 3, 2014
Posted
August 12, 2014
Terminated
January 30, 2017
Recall Number
Z-2184-2014
Quantity
9,498 devices total
Firm Location
Raynham, MA
Official Source
View on FDA website ↗

Reason for Recall

To clarify the CODMAN CERTAS Valve virtual off (setting 8) pressure specification outlined in the product's Instructions for Use (IFU).

Distribution

Worldwide Distribution - USA including AL, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV, WY; and Internationally to: Canada, Australia, Austria, Belgium, Brazil, Czech Republic, Denmark, Egypt, Finland, France, Germany, Great Britain, Greece, Hungary, Ireland, Israel, Italy, Kuwait, Lebanon, Luxembourg, Netherlands, Norway, Portugal, Russia, Saudi Arabia, Slovak Republic, Slovenia, South Africa, Spain, Sweden, Switzerland, Turkey.

Lot / Code Info

Lot Numbers: CLNCB1, CMBBP1, CMBCPN, CMCBWG, CMCC2V, CMDCN8, CMDDJN, CMGB8H, CMI

Root Cause

Device Design

Action Taken

Codman Neuro sent an Urgent Medical Device Safety Notification dated July 17, 2014, to clinicians who use the device. The notification clarifies the CODMAN CERTAS Valve virtual off (setting 8) pressure specification outlined in the product's Instructions for Use (IFU). Consignees are also asked to complete the enclosed Acknowledgement Form and fax the completed form to 1-888-239-1305. For questions or concerns regarding this notification, consignees are instructed to contact their local Codman Neuro Representative or Scientific and Medical Affairs at SciMedAffairs@its.jnj.com or (866) 685-7325. For questions regarding this recall call 508-828-2726.