Ambu, Aura Gain - Single Use Laryngeal Mask- Sterile for use as an alternative to a face mask for achieving and maintaining control of the airway during routine emergency anesthetic procedures.

Company
Ambu Inc.
Recall Initiated
February 10, 2016
Posted
March 24, 2016
Terminated
June 22, 2016
Recall Number
Z-1236-2016
Quantity
Total Numbers of unit distributed: 45990 pes (USA)
Firm Location
Columbia, MD

Reason for Recall

Ambu, Aura Gain laryngeal Mask, was found to have lack of Sterility Assurance. The firm discovered the problem during final quality control at manufacturing site. Damage to the sterile barrier (the pouch) may compromise a medical device's sterility,leading to contamination of the product and,ultimately,patient complications.

Distribution

Please see attached consignee list

Lot / Code Info

Catalog Number:408300000  Size #3:All lots from 1632796 through lot number 1808911except lot number 1779966,1782239, 1795914   Catalog Number: 408400000 Size #4 All lots from 1632799 through lot number 1790075.   Catalog Number:408500000 Size #5 All lots from 1632802 through lot number 1810863.

Root Cause

Process control

Action Taken

Filed Safety Notice was sent via USPS Certified Mail to all customers on Dec 10,2015.