Lifesaver Single Patient Use Manual Resuscitator Product Usage: The Hudson RCI Lifesaver Single Patient Use manual resuscitator with pressure monitoring port is a disposable medical device intended for use on patients requiring temporary augmentation of ventilation with or without supplemental oxygen delivery during episodes of acute ventilatory failure or insufficiency.

This recall has been terminated (originally issued June 24, 2015).

Company
Teleflex Medical
Recall Initiated
May 14, 2015
Posted
June 24, 2015
Terminated
May 13, 2024
Recall Number
Z-1809-2015
Quantity
9,333 units
Firm Location
Research Triangle Park, NC
Official Source
View on FDA website ↗

Reason for Recall

The intake port may be blocked which can cause the bag to fail to fill.

Distribution

Worldwide Distribution - US Nationwide in the states of AL, AK, AZ, AR, CA, CO, CT, FL, GA, IL, IA, KS, LA, MA, MI, MN, MS, MO, NE, NH, NY, NC, OH, OK, PA, RI, SD, TN, TX, VA, WA, including Puerto Rico and the countries of Australia, Bahamas, Canada, Guatemala, and Mexico.

Lot / Code Info

Product Code 5361 - 6 Digit Lot No. 140514 - 4 Digit Lot No. 1420; 6 Digit Lot No. 140616 4 Digit Lot No. 1425, 6 Digit Lot No. 140617 4 Digit Lot No. 1425; 6 Digit Lot No. 140621 4 Digit Lot No. 1425; 6 Digit Lot No. 140806 4 Digit Lot No. 1432; 6 Digit Lot No. 141103 4 Digit Lot No. 1445; 6 Digit Lot 141117 - No. 4 Digit Lot No. 1447; 6 Digit Lot No. 141201- 4 Digit Lot No. 1449; 6 Digit Lot No. 141224 - 4 Digit Lot No. 1452; 6 Digit Lot No. 141227 - 4 Digit Lot No. 1452; 6 Digit Lot No. 150112 - 4 Digit Lot No. 1503; Product Code 5362 - 6 Digit Lot No. 140504 - 4 Digit Lot No. 1419; 6 Digit Lot No. 140515 - 4 Digit Lot No. 1420; 6 Digit Lot No. 140517 - 4 Digit Lot No. 1420; 6 Digit Lot No. 140617 - 4 Digit Lot No. 1425; 6 Digit Lot No. 140621 - 4 Digit Lot 1425; No. 6 Digit Lot No. 140812 - 4 Digit Lot No. 1433; 6 Digit Lot No. 140819 - 4 Digit Lot No. 1434; 6 Digit Lot No. 140823 - 4 Digit Lot No. 1434; 6 Digit Lot No. 140929 - 4 Digit Lot No. 1440; 6 Digit Lot No. 141110 - 4 Digit Lot No. 1446; 6 Digit Lot No. 141115 - 4 Digit Lot No. 1446; 6 Digit Lot No. 141201 - 4 Digit Lot No. 1449; 6 Digit Lot No. 141214 - 4 Digit Lot No.1451; 6 Digit Lot No. 150112 4 Digit Lot No. 1503; Product Code 5364 6 Digit Lot No. 140507 - 4 Digit Lot No. 1419; 6 Digit Lot No. 140508 4 Digit Lot No. 1419; 6 Digit Lot No. 140514 - 4 Digit Lot No. 1420; 6 Digit Lot No. 140608 - 4 Digit Lot No. 1424; 6 Digit Lot No. 140614 - 4 Digit Lot No. 1424; 6 Digit Lot No. 140623 - 4 Digit Lot No. 1426; 6 Digit Lot No. 14628 - 4 Digit Lot 1426 - No. 6 Digit Lot No. 140714 - 4 Digit Lot No. 1429; 6 Digit Lot No. 140719 - 4 Digit Lot No. 1429; 6 Digit Lot No. 14806 - 4 Digit Lot No. 1432; 6 Digit Lot No. 140809 4 Digit Lot No. 1432; Product Code 5466 6 Digit Lot No. 140520 - 4 Digit Lot No. 1421; 6 Digit Lot No. 140923 - 4 Digit Lot No. 1439; 6 Digit Lot No. 140924 - 4 Digit Lot No. 1439; Product Code 45362 - 6 Digit Lot No. 140317 - 4 Digit Lot No. 1412; 6 Digit Lot No. 140630 - 4 Digit Lot No. 1427; 6 Digit Lot No. 141013 - 4 Digit Lot No. 1442.

Root Cause

Nonconforming Material/Component

Action Taken

Teleflex Medical sent an Urgent Medical Device Recall Notification letter dated May 14, 2015, to all consignees asking them to immediately discontinue distribution and quarantine the recalled product. The letter also requested a sub-recall of the product. The recall letter also included a response form which is to be returned to Teleflex by emailing it to recall@teleflex.com or fax it to 1-855-419-8507, Attn: Customer Service. Customers with questions were instructed to contact their local sales representative or Customer Service at 1-866-246-6990. For questions regarding this recall call 610-378-0131.