Passport V Monitor, Multiparameter Patient Monitor(with Arrhythmia Detection and Alarms) Intended for intra hospital use under the direct supervision of a licensed healthcare practitioner.

Company
Mindray Ds Usa, Inc. D.B.A. Mindray North America
Recall Initiated
January 23, 2015
Posted
March 16, 2015
Terminated
April 20, 2017
Recall Number
Z-1280-2015
Quantity
519 units
Firm Location
Mahwah, NJ

Reason for Recall

An issue has been identified with Passport V Monitors invasive blood pressure function (IBP) which may provide an incorrect IBP measurement

Distribution

Nationwide Distribution

Lot / Code Info

6100F- PA00195 6100F- PA00197 6100F- PA00219 6100F- PA00221 6100F- PA00267 6100F- PA00269 6100F- PA00291 6100F- PA00293

Root Cause

Process control

Action Taken

Mindray DS USA Inc sent Urgent: Passport V Monitor Field Corrective Action letters dated 1/23/2015 to hospital administrators. The letters informed customers that the IBP module may provide an incorrect IBP measurement. The letter also explained the adverse effects on patients. Customers are asked to either calibrate the IBP following the instructions provided; or, contact a Mindray Service Representative, who can perform the calibration, at 1-800-288-2121. In the case of the second option, it is recommended that the monitor be removed from service until the calibration is performed. If the first option is chosen, customers are to notify a Service Representative upon completion of the action. Customers can direct their questions to Diane Arpino at 201-995-8407.