Restoris Partial Knee Application (PKA) RIO (TGS 2.0). For use with the Robotic Arm Interactive Orthopedic System (RIO), is indicated for use in surgical knee procedures.

This recall has been terminated (originally issued February 19, 2015).

Company
Mako Surgical Corporation
Recall Initiated
January 22, 2015
Posted
February 19, 2015
Terminated
December 14, 2016
Recall Number
Z-1128-2015
Quantity
228 units
Firm Location
Plantation, FL
Official Source
View on FDA website ↗

Reason for Recall

When using the MAKOplasty partial knee Arthroplasty application, the burr continues spinning outside of the stereotactic boundary and after the control switches (foot pedal and trigger) cease to be activated

Distribution

Nationwide Distribution-including the states of AK, AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, LA, MA, MD, MI, MO, MS, NC, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WV and Hawaii.

Lot / Code Info

All lots, Part Numbers: 201000, 203999, 207300, and 209930.

Root Cause

Device Design

Action Taken

On January 22, 2015 the firm sent an IMPORTANT MEDICAL DEVICE CORRECTION asking consignees to complete the attached acknowledgement form and return them by email, fax or mail. The notice also noted the RIO system can continue to be used. The likelihood of occurrence is rare (.035% of cases) and the event is readily detectable and can be resolved. Should you have any questions regarding this Important Medical Device Recall Notice, to contact us at (954) 628-1721.