RecallDepth

MK5 NX Controller--Wheelchair controller

Company
Invacare Corporation
Recall Initiated
June 24, 2014
Posted
July 22, 2014
Terminated
October 15, 2014
Recall Number
Z-2091-2014
Quantity
819
Firm Location
Elyria, OH

Reason for Recall

M41 Powered Wheelchairs received through Invacare Service Parts (ISP) have controllers that were programmed in a way this is not aligned to standard factory settings. The controller during different settings such as max forward speed, max forward acceleration, max reverse speed and max turning speed can result in momentary loss of control.

Distribution

Nationwide Distriution: CA, SC, IL, UT, TN, KS, HI, NJ, LA, TX,FL, IN, AL, AZ, MN, WI, MN, MI, AR,MA, GA, MS, NJ, NM, OK, ME, PA, NY, NC, OK, MO, WA, IA, OR,KY, MD, NE, ID, VA, WA, ND, MT, OH, NH, CT. Also shipped to Canada

Lot / Code Info

P/N 1143839

Root Cause

Process control

Action Taken

The controllers will be reporogrammed in the field upon installation Invacare Sent out a Technical Services Bulletin (Oct. 15, 2012) . Please call 800-832-4707, Monday-Friday, 8:00 am- 7:00 pm EST, for assistance. If you are aware of a consumer complaint regarding this controller, please contact the Invacare Customer Service Complaints Department at 877-374-8919, Monday-Friday, 8:00am- 7:00pm or cscomplaints@invacare.com

More recalls by Invacare Corporation

View all recalls by this company →